HPLC Determination of Vitamin A and E
Introduction
Following GB method for Vitamin A and Vitamin E as it is written with a 250 x 4.6 mm Ascentis® Express C30 column gives baseline resolution between the critical pair (gamma-, and beta-tocopherol; peaks 3 and 4). When a 150 x 4.6 mm Ascentis® Express C30 column is used, baseline resolution is still maintained with a separation that is complete in less than half of the time of the original method. Additionally, the flow rate can be doubled with the 250 x 4.6 mm Ascentis® Express C30 column, which gives a separation that is less than ten minutes compared to the original method. This application demonstrates how converting methods to superficially porous particle technology improves throughput.

Vitamin A or retinyl acetate

General structure of tocopherol
Tocopherol | R1 | R2 |
Alpha (α) | CH₃ | CH₃ |
Beta (β) | CH₃ | H |
Gamma (γ) | H | CH₃ |
Delta (δ) | H | H |
Experimental Conditions | |
---|---|
Column | Ascentis® Express C30 (2.7µm) 250 x 4.6 mm |
Detection | UV @ 294 nm (semi-micro flow cell; 5 µL) |
Injection volume | 10 µL |
Flow rate | 0.8 mL/min |
Temperature | Column: 20 °C Autosampler: ambient |
Pressure drop | 237 bar (3437 psi) |
Mobile Phase | [A] = Water, [B] = Methanol |
Gradient | 96% B until 13 min, 96 - 100% B in 7 min, hold at 100% B for 4 min, 100 - 96% B in 0.5 min, hold at 96% B for 5.5 min |
Diluent | Methanol:Ethanol:Hexane 6:3:1 (v/v) |
Standard Solution | 20 µg/mL retinyl acetate, 400 µg/mL δ-tocopherol, 100 µg/mL γ-tocopherol, 1 mg/mL β-tocopherol, and 300 µg/mL α-tocopherol in diluent |

Experimental Conditions | |
---|---|
Column | Ascentis® Express C30 (2.7µm) 150 x 4.6 mm |
Detection | UV @ 294 nm (semi-micro flow cell; 5 µL) |
Injection Volume | 10 µL |
Flow Rate | 0.8 mL/min |
Temperature | Column: 20 °C Autosampler: ambient |
Pressure Drop | 153 bar (2219 psi) |
Mobile Phase | [A] = Water, [B] = Methanol |
Gradient | 96% B until 13 min, 96 - 100% B in 7 min, hold at 100% B for 4 min, 100 - 96% B in 0.5 min, hold at 96% B for 5.5 min |
Diluent | Methanol:Ethanol:Hexane 6:3:1 (v/v) |
Standard Solution | 20 µg/mL retinyl acetate, 400 µg/mL δ-tocopherol, 100 µg/mL γ-tocopherol, 1 mg/mL β-tocopherol, and 300 µg/mL α-tocopherol in diluent |

Experimental Conditions | |
---|---|
Column | Ascentis® Express C30 (2.7µm) 150 x 4.6 mm |
Detection | UV @ 294 nm (semi-micro flow cell; 5 µL) |
Injection Volume | 10 µL |
Flow Rate | 0.8 or 1.6 mL/min |
Temperature | Column: 20 °C Autosampler: ambient |
Pressure Drop | 243 bar (3524 psi) @ 0.8 mL/min 479 bar (6946 psi) @ 1.6 mL/min |
Mobile Phase | [A] = Water, [B] = Methanol |
Gradient | 96% B until 13 min, 96 - 100% B in 7 min, hold at 100% B for 4 min, 100 - 96% B in 0.5 min, hold at 96% B for 5.5 min |
Diluent | Methanol:Ethanol:Hexane 6:3:1 (v/v) |
Standard Solution | 20 µg/mL retinyl acetate, 400 µg/mL δ-tocopherol, 100 µg/mL γ-tocopherol, 1 mg/mL β-tocopherol, and 300 µg/mL α-tocopherol in diluent |


To continue reading please sign in or create an account.
Don't Have An Account?