Skip to Content
MilliporeSigma

GC Analysis of Organochlorine Pesticides in Water on SPB®-5 after SPME using 100 μm PDMS Fiber

GC Analysis of Organochlorine Pesticides in Water on SPB®-5 after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

SPME fiber

Product No.
Description
Pricing

SPME fiber assembly Polydimethylsiloxane (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

CONDITIONS

sample/matrix

200 ppt each in 2 mL water

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

15 min immersed in water with rapid stirring

column

SPB-5, 15 m × 0.20 mm I.D., 0.20 μm (24165-U)

oven

120 °C (1 min) to 180 °C at 30 °C/min, then to 290 °C at 10 °C/min

inj. temp.

260 °C (splitless - closed 3 min)

detector

ECD, 300 °C

carrier gas

helium, 37 cm/sec (set at 120 °C)