콘텐츠로 건너뛰기
Merck

US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap

US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap suitable for GC

Materials

관련 제품

제품 번호
설명
가격

Purge/Trap K Vocarb® 3000

for use with OI Analytical Eclipse 4660, 4560

분석 컬럼

제품 번호
설명
가격

SPB®-624 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 1.40 μm

표준

제품 번호
설명
가격

EPA 524.2 Rev 4 Update Mix

2000 μg/mL in methanol, analytical standard

EPA 524 Internal Standard Mix

certified reference material, 2000 μg/mL each component in methanol

EPA 524.2 Rev 4 Update Ketones Mix

certified reference material, 2000 μg/mL in methanol: water (9:1)

EPA 502/524 Volatiles Organic Calibration Mix A (without gases)

certified reference material, 200 μg/mL each component in methanol, ampule of 1 mL

Volatile Organic Compounds Mix 6

certified reference material, 2000 μg/mL each component in methanol

CONDITIONS

sample/matrix

each analyte at 10 ppb in 25 mL water

purge trap

VOCARB 3000 "K" (24940-U)

dry purge

2 min

desorption preheat

205 °C

desorption process

40 mL/min at 210 °C for 2 min

bake

260 °C for 10 min

transfer line/valve temp.

110 °C

sample preparation

purge is 40 mL/min at 25 °C for 11 min

column

SPB-624, 30 m x 0.25 mm I.D., 1.4 μm (24255)

oven

40 °C (2 min), 7 °C/min to 135 °C, 30 °C/min to 230 °C (3 min)

carrier gas

helium, 1.2 mL/min

injection

30:1 split

liner

0.75 mm I.D. SPME

설명

분석 메모

Method 524.2 describes the analysis of volatile organic compounds in surface, ground, and drinking water. This application illustrates the analysis of the compounds listed in this method on a 30m x 0.25 mm I.D., 1.4 μm SPB-624 column using purge and trap / GC-MS.