Skip to Content
MilliporeSigma

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V (Method 2)

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V (Method 2) application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIROBIOTIC® V Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

Application

Product No.
Description
Pricing

HPLC Analysis of Warfarin Anticoagulant on Ascentis® C18

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® CYCLOBOND I 2000 (UV at 254 nm)

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® CHIROBIOTIC® V2

application for HPLC

HPLC Analysis of Warfarin Enantiomers on Astec® Cellulose DMP

application for HPLC

standard

Product No.
Description
Pricing

Warfarin

analytical standard

Warfarin

PESTANAL®, analytical standard

(R)-(+)-Warfarin

≥97% (HPLC)

(S)-(−)-Warfarin

≥97% (HPLC)

CONDITIONS

column

CHIROBIOTIC V, 25 cm x 4.6 mm I.D., 5 μm particles (11024AST)

mobile phase

[A] 0.1% triethylamine acetate, pH 5.0; [B] THF; (80:20, A:B)

flow rate

1 mL/min

column temp.

ambient

detector

UV, 254 nm

injection

1 μL

sample

4 mg/mL in methanol

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC