추천 제품
material
glass support
silica gel 60 matrix
Quality Level
특징
binder Organic Polymer
fluorescent indicator
포장
pkg of 25 plates
기술
HPTLC: suitable
레이어 두께
200 μm
플레이트 길이 × 너비
20 cm × 10 cm
입자 크기
5-6 μm
공극 크기
60 Å medium pore diameter
저장 온도
2-30°C
일반 설명
25 Glass plates 20 x 10 cm
Our NH2, CN and DIOL -modified silica sorbents are less polar then conventional silica phases, making them ideal for separating hydrophilic or charged substances. Our moderately polar cyano- and diol-modified silica plates can be used for both normal phase and reversed phase systems. An alternative to PEI cellulose, amino-modified NH2 plates provide weak basic ion exchange characteristics with special selectivity for charged compounds.
분석 메모
Specific surface area (according to BET; 5-Pt. measurement): 480 - 540 m²/g
Pore volume (N₂-isotherm): 0.74 - 0.84 ml/g
d 50 (laser diffraction, size distribution): 5 - 7 µm
Layer thickness: 150 - 200 µm
Deviation of layer thickness per plate: ≤ 35 µm
Chromatographic testings:
A) Steroidtest, straight phase
hRf-values
- hydrocortisone, straight phase:
4 - 8
- Reichstein S, straight phase: 11 - 18
- Methyltestosterone, straight phase: 27 - 36
B) Steroid test, reversed phase
- methyltestosterone, steroid test: 48 - 68
- Reichstein S, steroid test: 58 - 78
- hydrocortisone, steroid test: 68 - 88
Typical value determined on an activated plate (120°C, 15 min)
Eluent a) petroleum benzine 40-60°/acetone (80/20 v/v)
Eluent b) acetone/water (30/70 v/v)
Pore volume (N₂-isotherm): 0.74 - 0.84 ml/g
d 50 (laser diffraction, size distribution): 5 - 7 µm
Layer thickness: 150 - 200 µm
Deviation of layer thickness per plate: ≤ 35 µm
Chromatographic testings:
A) Steroidtest, straight phase
hRf-values
- hydrocortisone, straight phase:
4 - 8
- Reichstein S, straight phase: 11 - 18
- Methyltestosterone, straight phase: 27 - 36
B) Steroid test, reversed phase
- methyltestosterone, steroid test: 48 - 68
- Reichstein S, steroid test: 58 - 78
- hydrocortisone, steroid test: 68 - 88
Typical value determined on an activated plate (120°C, 15 min)
Eluent a) petroleum benzine 40-60°/acetone (80/20 v/v)
Eluent b) acetone/water (30/70 v/v)