Skip to Content
MilliporeSigma

GC Analysis of C1-C5 Hydrocarbons on Alumina Sulfate PLOT

GC Analysis of C1-C5 Hydrocarbons on Alumina Sulfate PLOT suitable for GC

CONDITIONS

column

Alumina sulfate PLOT, 50 m x 0.53 mm I.D. (28324-U)

oven

35 °C (2.5 min), 5 °C/min to 150 °C (4.5 min)

column temp.

200 °C

detector

FID, 230 °C

carrier gas

helium, 3.0 mL/min

injection

10 μL direct using a sampling valve, 150 °C

sample

18-component C1-C5 hydrocarbon standard in nitrogen, each analyte at 25 ng on-column