ABSTRACT

Limits of detection are important issues for GC/MS screening for anabolic agents and for confirmation of various drugs of abuse. We compared a quadrupole ion trap (QIT) operated in two different selected-ion storage modes and a quadrupole mass filter (QMF) operated in the selected-ion monitoring mode. Results with the model compound tetrachlorobenzene indicate that, for simultaneous monitoring of more than four ions, the QIT operated in a frequency-modulated selected-ion storage mode has better limits of detection than the QMF. Use of a single-ion storage technique gave results similar to those of the QMF. We also evaluated both QIT selected-ion storage approaches for the limits of detection of the trimethylsilyl derivatives of four anabolic steroid metabolites and the beta-agonist clenbuterol. We found no improvement in detection limits over that of a similar method with selected-ion monitoring and a QMF when four anabolic steroid metabolites and clenbuterol were extracted from a urine matrix. The lack of improvement in the limit of detection resulted from matrix background signals at masses similar to those of the steroids.

MATERIALS
Product Number
Brand
Product Description

Sigma-Aldrich
1,2,3,4-Tetrachlorobenzene, 98%