Ascentis® Express U/HPLC Column Selection

Column selection by phase chemistry

The column selectivity has the highest influence on resolution in chromatography. Therefore, the selection of the best suitable column chemistry for the target analytes is an important selection parameter. C18 column chemistries are typically the first choice. Nevertheless, when a C18 doesn't give the desired separation or the sample contains compounds that are known to be difficult to retain or resolve on a C18, consider changing stationary phases. The range of selectivity provided by Ascentis® Express makes this easy. The flowchart below helps guide users in the selection of an Ascentis® Express phase, based on the particular compound type or separation challenge

Ascentis® Express HPLC Column Selection Guide

column selection by column dimension

The separation goals, detection method and the instrument used for the separation define the column dimension and the particle size.

Select Column ID

NeedsColumn ID (mm)Typical Flow Rate (mL/min)
Best Suitability for Mass Spec and UHPLC2.10.2 – 1.0
Solvent Savings in HPLC3.00.5 – 2.0
Standard HPLC4.61.0 – 3.0

Select Column Length

NeedsColumn Length (mm)
Maximize Speed20 - 75
Balance of Resolution and Speed100
Maximize Resolution150

Select Particle Size

NeedsParticle Size (µm)
UHPLC / High Throughput2.0
Fast on any HPLC System2.7
"Plug and Play" for Improving Existing HPLC Methods5.0