コンテンツへスキップ
Merck

G005538

GCxGC Analysis of FAMEs in Human Plasma on Equity-1 x SUPELCOWAX 10, Non-Polar to Polar Strategy

suitable for GC

ログイン組織・契約価格を表示する


About This Item

テクニック

gas chromatography (GC): suitable

試験パラメータ

column: (1D) Equity 1, 30 m x 0.25 mm I.D., 0.25 μm (28406-U)
column: (2D) SUPELCOWAX 10, 0.95 m x 0.10 mm I.D., 0.10 μm (24343)
oven: 180 °C, 3 °C/min to 280 °C (10 min)
oven: (modulator) LMCS Everest longitudinally modulated cryogenic system, modulation period 6 sec
inj. temp.: 280 °C
detector: FID, 300 °C, 125Hz; MS, 20000 amu/sec
carrier gas: hydrogen
sample: human plasma

アプリケーション

clinical

詳細

The 0.95 m SUPELCOWAX 10 column was obtained by cutting lengths from a standard length column (e.g. 5, 10, 15 m).

アナリシスノート

Sample prep: 100 mL of plasma was saponified with 1 mL of methanolic sodium methoxide (0.5% w/v) at a temperature of 100 °C for 15 min in a closed Pyrex tube. The subsequent methylic esterification was achieved with 1 mL boron trifluoride-methanol complex at a temperature of 100 °C for 15 min. The fatty acid methyl esters (FAMEs) were extracted by adding 1 mL of n-hexane to the mixture. One mL of a saturated NaCl solution was added to the entire mixture, which was then agitated manually for 2 min, followed by centrifugation at 3000 rpm for 10 min. After, anhydrous Na2SO4 was added to the Pyrex tube for drying and 0.9 mL of the n-hexane FAME layer was transferred to a GC injector vial. A 0.1 mL volume of 1000 ppm internal standard (C13:0) was also added to the GC vial.

法的情報

分析カラム


最新バージョンのいずれかを選択してください:

試験成績書(COA)

Lot/Batch Number

申し訳ございませんが、現在この製品のCOAをオンラインで入手できません。

サポートが必要な場合は、お問い合わせください カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)