콘텐츠로 건너뛰기
Merck
모든 사진(1)

문서

G005826

LC/MS (TOF) Analysis of Dermorphin on Ascentis® Express F5

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Ascentis Express F5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53569-U)
mobile phase: 2 mM ammonium formate (90:10 acetonitrile:water) adjusted to pH 4.5 with formic acid
flow rate: 0.4 mL/min
column temp.: 35 °C
detector: ESI(+), TIC 100 – 1000 m/z; APCI (+), XIC m/z = 150.1, 203.7, 378.2
injection: 1 μL
sample: 10 μg/mL in 75:25 acetonitrile:water

적합성

application for HPLC

응용 분야

forensics and toxicology

분석 메모

Dermorphin is a heptapeptide derived from frogs. It is a painkiller and used illicitly in racehorses so they continue to run through the pain. A rapid analysis of dermorphin standard is shown here on an Ascentis Express F5 column. The HILIC mode separation provided the necessary retention and MS-compatibility, while the pentafluorophenyl (F5) phase chemistry elimated the chelating interactions that are problematic on bare silica.

법적 정보

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Dermorphin trifluoroacetate salt, ≥98% (HPLC)

가장 최신 버전 중 하나를 선택하세요:

시험 성적서(COA)

Lot/Batch Number

죄송합니다. 지금은 이 제품에 대한 COA이(가) 온라인에서 제공되지 않습니다.

도움이 필요하시면 연락하세요. 고객 지원 부서

이 제품을 이미 가지고 계십니까?

문서 라이브러리에서 최근에 구매한 제품에 대한 문서를 찾아보세요.

문서 라이브러리 방문

자사의 과학자팀은 생명 과학, 재료 과학, 화학 합성, 크로마토그래피, 분석 및 기타 많은 영역을 포함한 모든 과학 분야에 경험이 있습니다..