HPLC Analysis of Insulin (Species Variants) on BIOshell A160 Peptide C18

application for HPLC

Sign Into View Organizational & Contract Pricing


About This Item

technique(s)

HPLC: suitable

test parameters

column: BIOshell A160 Peptide C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (66905-U)
mobile phase: [A] 70:30, water with 0.1% TFA:acetonitrile with 0.09% TFA; [B] 55:45, water with 0.1% TFA:acetonitrile with 0.09% TFA
gradient: 0 to 100% B in 7.5 min
flow rate: 0.3 mL/min
pressure: 4040 psi (278 bar)
column temp.: 30 °C
detector: UV, 215 nm
injection: 0.5 μL
sample: 200 μg/mL insulin (bovine, human, porcine) in 0.1% formic acid

suitability

application for HPLC

application(s)

life science and biopharma
pharmaceutical (small molecule)

Analysis Note

These three insulins have more than 76% identical positions in a primary structure alignment. The human and porcine proteins are most similar with 85% identity. Baseline resolution of these insulins in achieved on BIOshell A160 Peptide C18, with relative band spacing of these three peaks approximating their relative sequence similarities. BIOshell A160 Peptide C18 is ideally suited for RP chromatography of small proteins or large peptides.

Legal Information

Analyte

Description

    Insulin from bovine pancreas, ≥25 USP units/mg (HPLC), powder
    Insulin human, powder, recombinant, expressed in E. coli, suitable for cell culture
    Insulin from porcine pancreas, powder, ≥27 USP units/mg

Choose from one of the most recent versions:

Certificates of Analysis (COA)

Lot/Batch Number

Sorry, we don't have COAs for this product available online at this time.

If you need assistance, please contact Customer Support.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Our team of scientists has experience in all areas of research including Life Science, Material Science, Chemical Synthesis, Chromatography, Analytical and many others.

Contact Technical Service