Skip to Content
MilliporeSigma

HPLC Analysis of Methadone Enantiomers on Astec® CHIROBIOTIC® V2

HPLC Analysis of Methadone Enantiomers on Astec® CHIROBIOTIC® V2 application for HPLC

Materials

analytical column

Product No.
Description
Pricing

Astec® CHIROBIOTIC® V2 Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

standard

Product No.
Description
Pricing

(±)-Methadone hydrochloride

powder, ≥98%

DL-Methadone-d3 solution

98 atom % D, 1 mg/mL in methanol, drug standard

CONDITIONS

column

CHIROBIOTIC V2, 25 cm x 4.6 mm I.D., 5 μm particles (15024AST)

mobile phase

[A] methanol; [B] 20 mM ammonium formate; (95:5, A:B)

flow rate

1 mL/min

column temp.

20 °C

detector

UV, 205 nm

injection

2 μL

sample

methadone, 1 mg/mL in mobile phase

Description

Legal Information

Astec is a registered trademark of Merck KGaA, Darmstadt, Germany
CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC