UHPLC Analysis of Bisphenol A and Analogous Compounds in Infant Formula on Titan C18 with Fluorescence Detection after SPE using SupelMIP® Bisphenol A

UHPLC Analysis of Bisphenol A and Analogous Compounds in Infant Formula on Titan C18 with Fluorescence Detection after SPE using SupelMIP® Bisphenol A application for UHPLC, application for SPE

Materials

SPE tube or plate

Product No.
Description
Pricing

SupelMIP® SPE - Bisphenol A

bed weight 100 mg, volume 6 mL, pk of 50

Standard

Product No.
Description
Pricing

Bisphenol A diglycidyl ether

Bis(4-hydroxyphenyl)methane

98%

Bisphenol A

≥99%

Bisphenol C

analytical standard

Bisphenol G

analytical standard

accessory

Product No.
Description
Pricing

Screw cap (open-top), 9 mm thread, with liner

clear cap, PTFE/silicone septum, pkg of 100 ea

ABC Vial, 2 mL (vial only), 9 mm thread

clear glass vial (with marking spot), pkg of 100 ea

sample prep

Product No.
Description
Pricing

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

CONDITIONS

sample/matrix

liquid infant formula

SPE well plate

SupelMIP Bisphenol A, 100 mg/3 mL (52775-U)

condition

3 mL 2% formic acid in methanol, 3 mL acetonitrile, 3 mL water

sample addition

as prepared

washing

9 mL of water, 6 mL of acetonitrile:water (40:60); vacuum was pulled for 3 min after the last wash

eluate post-treatment

Sample was evaporated to dryness at 45 °C, reconstuituted in 0.5 mL of acetonitrile:water (40:60). All samples were filtered through Durapore membrane filter units prior to analysis.

column

Titan C18, 10 cm x 2.1 mm I.D., 1.9 μm particles (577124-U)

mobile phase

[A] water; [B] acetonitrile

gradient

40 to 80% B in 2.5 min; held for 4.5 min; to 100%B in 0.1 min; held for 2.4 min; to 40% B in 0.1 min; held for 2.9 min

flow rate

0.25 mL/min for 7 min; to 0.40 mL/min in 0.1 min; held for 2.4 min; to 0.25 mL/min in 0.1 min.; held for 2.9 min

pressure

3916 psi (270 bar)

column temp.

35 °C

detector

fluorescence, ex: 230 nm, em: 315 nm

injection

5 μL

elution

3 mL of methanol:acetonitrile (75:25) (A017)

Description

Analysis Note

BPA can leach into foods from the contact between food and the packaging. Structurally analogous compounds could also be used in place of BPA during package manufacturing as BPA comes under more stringent regulations. The presented method was developed for analysis of this class of compounds and applied to infant formula analysis. The use of molecularly imprinted polymers during cleanup produced samples that can be analysed by fluoresent detection.

Other Notes

Sample Prep: 5 mL of liquid infant formula was spiked with 10 ppb of analytes, mixed with 5.5 mL of acetonitrile, centrifuged at 4500 rpm for 10 minutes, top 3 mL were diluted with 3 mL water to load into SPE

Legal Information

SupelMIP is a registered trademark of Sigma-Aldrich Co. LLC
null