テクニック
HPLC: suitable
試験パラメータ
column: Ascentis Express OH5, 10 cm x 2.1 mm I.D., 2.7 μm particles (53757-U)
mobile phase: [A] 5 mM ammonium acetate, pH 5.0 with acetic acid in 95:5, acetonitrile:water; [B] 5 mM ammonium acetate, pH 5.0 with acetic acid in 80:20, acetonitrile:water
gradient: 0% B held for 1 min; to 100% B in 10 min; held at 100% B for 1 min
flow rate: 0.3 mL/min
column temp.: 25 °C
detector: UV, 250 nm
injection: 2 μL
sample: 10 - 100 μg/mL in 95:5, acetonitrile:water (pH of buffer stock (in water) was adjusted before further dilution with water and/or acentonitrile)
適合性
application for HPLC
アプリケーション
life science and biopharma
アナリシスノート
Nucleosides are involved in many aspects of metabolism. Analogs of nucleosides have medical application in antivirals or anticancer treatments. This application demonstrates the utility of hydrophilic interaction chromatography (HILIC) for the retention and resolution of nucleosides and analogs thereof. Elution is not solely explained by logD but by other ionic and solvophobic effects.
法的情報
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
アナライト
詳細
5-Methyluridine
DMT-2′O-TC-rU Phosphoramidite
2-Thiocytidine dihydrate
2′-O-Methylcytidine
Inosine
5-Methylcytidine
Cytidine
DMT-2′O-TC-rG(ib) Phosphoramidite
1-Methyladenosine
7-Methylguanosine
すべて表示 (10)
Standard
製品番号
詳細
価格
最新バージョンのいずれかを選択してください:
ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.
製品に関するお問い合わせはこちら(テクニカルサービス)