G1007494

UHPLC Analysis of Nucleosides on Purospher® STAR RP-18 endcapped

application for HPLC

로그인조직 및 계약 가격 보기


About This Item

기술

HPLC: suitable

검사 파라미터

column: Hibar Purospher® STAR RP-18 endcapped, 150 x 4.6 mm, 3 μm (1.50470.0001)
mobile phase: [A] 30 mM ammonium phosphate, pH 5.3 with ammonium hydroxide in water; [B] Methanol; 97.5 [A] / 2.5 [B]
flow rate: 1.0 mL/min
pressure: 3060 psi 211 Bar
column temp.: 25 °C
detector: UV 254
injection: 5 μL
sample: 5 μg/mL of 12 nucleosides in 98:2 30mM ammonium phosphate pH 5.3 with ammonium hydroxide in water: methanol

적합성

application for HPLC

응용 분야

life science and biopharma

분석 메모

Nucleosides are structural subunits of nucleic acids, the macromolecules that convey genetic information in living cells. They contain a molecule of sugar (pentose) bonded to a nitrogen-containing organic ring compound. The nitrogen-containing compound is either a pyrimidine (cytosine, thymine, or uracil) or a purine (adenine or guanine). Nucleosides, derivatives and / or analogues can be used in many different industries and applications as in agrochemistry (herbicides, fungicides and insecticides), therapeutic drugs including a range of antiviral (HIV, hepatitis C virus and herpes simplex) and even antibiotics.

법적 정보

Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany

분석물

설명

    Pseudouridine
    Cytidine
    3-methylcytidine methosulfate
    1-methyladenosine
    Uridine
    2-thiocytidine
    5-methylcytidine
    7-methylguanosine
    2′-O-methylcytidine
    Inosine
    Guanosine
    Ribothymidine
    모두 보기 (12)