コンテンツへスキップ
Merck

G006153

LC/MS (TOF) Analysis of Opioids and Opiate-Dependence Management Drugs on Ascentis® Express HILIC

application for HPLC

ログイン組織・契約価格を表示する


About This Item

テクニック

HPLC: suitable

試験パラメータ

column: Ascentis Express HILIC, 5 cm x 2.1 mm I.D., 2.7 μm particles (53934-U)
mobile phase: 2 mM ammonium formate in 90:10 (v:v) acetonitrile:water
flow rate: 0.3 mL/min
column temp.: 55 °C
detector: TOF/MS
injection: 2 μL
sample: 400 ng/mL in acetonitrile

適合性

application for HPLC

アプリケーション

clinical testing
forensics and toxicology

アナリシスノート

The rapid analysis by HPLC with TOF-MS detection of seven opioids is shown here using an Ascentis Express HILIC column. Cerilliant CRMs were used to ensure reliable MS identification and quantification.

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

アナライト

詳細

    Oxycodone solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    6-Acetylcodeine solution, 1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    6-Acetylmorphine, 1.0 mg/mL solution, 1 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®
    Codeine solution, 1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Hydrocodone solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Morphine solution, 1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®
    Hydromorphone solution, 1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®