コンテンツへスキップ
Merck

G1006927

LC/MS/MS Analysis of Pesticide Residues in Dates on the Ascentis® Express RP-Amide Column after QuEChERS Extraction

application for LC-MS, application for SPE

ログイン組織・契約価格を表示する


About This Item

テクニック

LC/MS: suitable
solid phase extraction (SPE): suitable

試験パラメータ

sample/matrix: 5 g ground dates spiked with pesticides at 10 ng/g
extraction process: add 10 mL water and 10 mL acetonitrile; shake 1 hr at 2500 rpm; add contents of Supel QuE citrate extraction tube (55227-U); . (shake for 1 minute; centrifuge at 5000 rpm fo 5 minutes and draw off supernatant)
clean-up tube: Supel QuE PSA/C18 (55288-U)
clean-up process: transfer 1 mL of the extract to 2 mLSupel QuE PSA/C18 tube; shake for 1 minute; centrifuge at 5000 rpm for 3 min; draw off supernatant for analysis.
sample preparation: Dilute 100 μL of cleaned extract to a final volume of 1 mL in 5 mM ammonium formate/0.1% formic acid in water; transfer diluted sample to vial for LC/MS/MS analysis
column: Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2 μm particles (51576-U)
mobile phase: (A) 5 mM ammonium formate with 0.1% formic acid in water; (B) 5 mM ammonium formate with 0.1% formic acid in 95:5 acetonitrile:H2O
gradient: held at 5% B from 0 to 1 min; to 100% B in 13 min; held at 100% B for 1.5 min; to 95% A in 0.5 min; held at 95% A for 1.5 min
flow rate: 0.4 mL/min
column temp.: 30 °C
detector: MS, ESI (+), MRM
injection: 5 μL

適合性

application for LC-MS
application for SPE

アプリケーション

food and beverages

アプリケーション

Freeze dates with liquid nitrogen and grind to a homogenous mixture.

アナリシスノート

The “quick, easy, cheap, effective, rugged, and safe” (QuEChERS) method can be used to analyze multiple pesticide residues found in dates. In this study, QuEChERS extraction and cleanup using PSA/C18 were used for the analysis of pesticide residues of interest in pistachios. Average recoveries obtained for replicate samples spiked at 10 ng/g were in the range of 70-120% for 23 of the 25 pesticides analyzed.

法的情報

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

アナライト

詳細

    Omethoate
    Butocarboximsulfoxide
    Aldicarb-sulfone
    Methomyl
    Carbendazim
    Trichlorfon
    Aldicarb
    Thiacloprid
    Thiophanate-methyl
    Carbofuran
    Methabenzthiazuron
    Chlorantraniliprole
    Spinetoram
    Malathion
    Bifenazate
    Triazophos isomer#1
    Triazophos isomer#2
    Rotenone
    Quinalphos
    Etrimfos
    Pyraflufen-ethyl
    MGK-264 #1
    MGK-264 #2
    Neburon
    Isoxathion
    Spiromesifen
    すべて表示 (26)

Standard

製品番号
詳細
価格

最新バージョンのいずれかを選択してください:

試験成績書(COA)

Lot/Batch Number

It looks like we've run into a problem, but you can still download Certificates of Analysis from our 資料 section.

サポートが必要な場合は、お問い合わせください カスタマーサポート

以前この製品を購入いただいたことがある場合

文書ライブラリで、最近購入した製品の文書を検索できます。

文書ライブラリにアクセスする

ライフサイエンス、有機合成、材料科学、クロマトグラフィー、分析など、あらゆる分野の研究に経験のあるメンバーがおります。.

製品に関するお問い合わせはこちら(テクニカルサービス)